Hospira Plum A+ 3

Hospira Plum A+ 3

The Plum A+®3 infusion system is designed to meet the growing demand for hospital wide device standardization, and serves a wide range of general floor and critical care needs. The infusion system consists of three component infusers, designated line 1, line 2, and line 3. By incorporating three lines into one unit, the Plum A+3 provides three primary lines, three secondary lines, and piggyback fluid delivery capabilities. Compatibility with the LifeCare® 5000 PlumSet® administration sets and accessories makes the Plum A+3 a convenient and cost-effective infusion system.

Features
  • Dose calculation
  • Loading dose
  • Multi-step programming
  • Drug list
  • Nurse call back
  • Delayed start setting
  • Air trap
  • Air removal/backpriming
  • Air detection (proximal and distal)
  • Serial communication
  • Variable rate cap
  • Alarm history
  • Nurse call relay connector
  • Volumes infused (A, B, total volumes)
  • Variable distal pressure settings
  • Non-pulsatile volumetric accuracy
  • Microprocessor control
  • Large LCD
  • Panel back illumination on mains power
  • Lockout switch
  • Standard fulfill, partfill, syringe, and vial use
  • Enteral and parenteral fluid delivery
  • Blood and blood product delivery
  • Wide range of standard and specialty administration sets
  • Password protected keypad lock

phone Call for price: (855) 477-8866

Download the operator's Manual

Description

Here are the specifications of  Hospira Plum A+ 3 Infusion Pump Repair  and hospira plum a 3 service manual.

Physical:
  • Dimensions: 19” H x 15” W x 14” D (excluding pole clamp and power cord storage)
  • Weight: 28 lbs (with batteries)
  • Casing: High-impact plastic
Electrical:
  • Power Requirements: 95-132 VAC; 47-62 Hz; 90 W
  • Power Cord: Hospital-grade AC cord; 10 feet; with transparent plug and retainer
  • Fuses: 0.5 A, 250 VAC
  • Battery: Three; sealed lead-acid; rechargeable; 6 V; internal
  • Battery Operation: A fully charged new battery provides approximately three hours of operation at 125 mL/hr or less, or delivers 250 mL if > 126 mL/hr
  • Recharge Time: Less than 6 hours (if operating at 125 mL/hr on one line)
  • Self-Discharge: 50% of charge is retained for a minimum of one month when the infuser is not connected to AC power or is not operating
Delivery Rate Ranges:
  • Delivery Rate (Lines A and B): 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments); 100 to 999 mL/hr (in 1 mL/hr increments)
  • Concurrent Delivery: 0.5 mL/hr minimum for each line
  • PlumSet: 500 mL/hr cumulative (A + B) maximum
  • VTBI Range: 0.1 to 99.9 mL (in 0.1 mL/hr increments); 100 to 9999 mL (in 1 mL/hr increments)
Alarms and Limits:
  • Distal Occlusion: The distal occlusion alarm sounds after the distal tubing or set outlet fitting becomes occluded
  • Proximal Occlusion: The proximal occlusion alarm sounds within two pumping cycles when the tubing proximal to the cassette becomes occluded
  • Distal Pressure Limit (without alarm): 1 to 15 psi; maximum pressure limit is user-selectable; factory setting is 6 psi
  • Maximum Infusion Pressure: 20 psi

Air-in-Line Alarm:

    • PlumSet (Distal): Bolus – 0.5 mL of air or larger; Cumulative – 0.5 mL of air out of 5.3 mL of fluid
    • PlumSet (Proximal): Bolus at 0.5 mL, total 1 mL (0.5 mL concurrent)
Alarms
    • Distal Occlusion
    • Proximal Occlusion
    • Distal air-in-line
    • Proximal air-in-line
    • Low battery
    • Door open while pumping
    • Lockout violation
    • VTBI complete
    • Valve/cassette test failure
    • Nurse call
    • No act

Download the operator’s Manual

https://web.facebook.com/infusionpumprepair/photos/a.101426038761773/102799898624387

Self-Discharge: 50% of charge is retained for a minimum of one month when the infuser is not connected to AC power or is not operating

Related Keyphrase:
  • Hospira Plum A+ 3 Infusion Pump
  • hospira plum a 3 service manual