Hospira Plum A+ with Mednet
The Plum A+® with Hospira MedNet® Software is an advanced medication management system designed to meet the growing demand for hospital wide, alternate site, and home healthcare device standardization. Advanced clinical capabilities, autoprogramming, networked communication, and a plug-and-play platform make the Plum A+ with Hospira MedNet Software a convenient and cost-effective multipurpose, multimode, flexible infusion system.
The host device contains a Communication Engine (CE) that provides wired Ethernet and wireless 802.11 a/b/g local area networking capabilities. Hospira MedNet networked application software is designed to allow a facility to customize and download a Drug Library for use with the infusion system.
Features
- Dose calculation
- Loading dose
- Multi-step programming
- Therapy selection
- Nurse call
- Delayed start setting
- Standby mode
- Drug library
- Piggyback/Concurrent delivery modes
- Titration
- Anti free-flow protection
- Networked communications
- Air removal/backpriming
- Air detection (proximal and distal)
- Alarm history
- Volumes infused (A,B, total volumes)
- KVO at dose end (1 mL/hr or less depending on delivery rate) or Continue Rate (CR) to continue at the current rate
- Variable distal pressure setting
- Nonpulsatile volumetric accuracy
- Microprocessor control
- Large LCD
- Panel back illumination on mains power
- Lockout switch
- Standard fulfill, partfill, syringe, and vial use
- Enteral and parenteral fluid delivery
- Blood and blood product delivery
Call for price:
(855) 477-8866
Description
Physical:
- Dimensions: 8” H x 8” W x 6” D (excluding pole clamp and power cord storage)
- Weight: 9.5 lbs (with battery)
- Casing: High-impact plastic
Electrical:
- Power Requirements: 120 VAC; 50-60 Hz; 35 W
- Power Cord: Hospital-grade AC cord; 10 feet; with transparent plug and retainer
- Fuses: 0.5 A, 250 VAC
- Battery: Sealed lead acid; 6 V; internal; rechargeable
- Battery Operation: A fully charged new battery provides approximately three hours of operation at 125 mL/hr, or delivers 250 mL if > 126 mL/hr
- Recharge Time: Less than 6 hours (if operating at 125 mL/hr on one line)
- The battery charges whenever the infuser is connected to AC power
Self-Discharge: 50% of charge is retained for a minimum of one month when the infuser is not connected to AC power or is not operating
Nurse Call System: Default: Normally-open (NO)
Delivery Rate Ranges:
- Delivery Rate (Lines A and B): 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments); 100 to 999 mL/hr (in 1 mL/hr increments)
- Concurrent Delivery: 0.5 mL/hr minimum for each line
- PlumSet: 500 mL/hr cumulative (A + B) maximum
- KVO: 1 mL/hr or the last primary delivery rate, whichever is less
- VTBI Range: 0.1 to 99.9 mL (in 0.1 mL/hr increments); 100 to 9999 mL (in 1 mL/hr increments)
- Accuracy: ± 5%
Alarms and Limits:
- Distal Occlusion: The distal occlusion alarm sounds after the distal tubing or set outlet fitting becomes occluded
- Proximal Occlusion: The proximal occlusion alarm sounds within two pumping cycles when the tubing proximal to the cassette becomes occluded
- Distal Pressure Limit
(without alarm): 1 to 15 psi; maximum pressure limit is user-selectable; factory setting is 6 psi - Maximum Infusion Pressure: 20 psi
Air-in-Line Alarm:
-
- PlumSet (Distal): Bolus – 0.5 mL of air or larger; Cumulative – 0.5 mL of air out of 5.3 mL of fluid
- PlumSet (Proximal): Bolus at 0.5 mL, total 1 mL (0.5 mL concurrent)
Communication:
Ethernet LAN: Shielded Ethernet cable plugged into an RJ-45 connector
Wireless LAN: Device name: Hospira MedNet Wireless 802.11 a/b/g Module
-
- Standards: IEEE 802.11 a/b/g
- Transmit Power: 802.11 b/g – 17.6 dBm; 802.11 a – 19 dBm
- Antenna: Integrated Surface Mount Antenna
- Certification: FCC Part 15.247, 15.407; IC RSS-210, RSS-102
- FCC ID: STJ-80411396001, IC: 5627A-80411396
Ethernet IP Address: 192.168.0.100
Subnet Mask: 255.255.0.0
Alarms
- Distal Occlusion
- Proximal Occlusion
- Distal air-in-line
- Proximal air-in-line
- Low battery
- Door open while pumping
- Lockout violation
- VTBI complete
- Valve/cassette test failure
- Nurse call
- No action alarm
- Infuser idle for two minutes