- Hospira Plum A+ with Mednet -

Hospira Plum A+ with Mednet

Hospira Plum A+ with Mednet

The Plum A+® with Hospira MedNet® Software is an advanced medication management system designed to meet the growing demand for hospital wide, alternate site, and home healthcare device standardization. Advanced clinical capabilities, autoprogramming, networked communication, and a plug-and-play platform make the Plum A+ with Hospira MedNet Software a convenient and cost-effective multipurpose, multimode, flexible infusion system.

The host device contains a Communication Engine (CE) that provides wired Ethernet and wireless 802.11 a/b/g local area networking capabilities. Hospira MedNet networked application software is designed to allow a facility to customize and download a Drug Library for use with the infusion system.

Features
  • Dose calculation
  • Loading dose
  • Multi-step programming
  • Therapy selection
  • Nurse call
  • Delayed start setting
  • Standby mode
  • Drug library
  • Piggyback/Concurrent delivery modes
  • Titration
  • Anti free-flow protection
  • Networked communications
  • Air removal/backpriming
  • Air detection (proximal and distal)
  • Alarm history
  • Volumes infused (A,B, total volumes)
  • KVO at dose end (1 mL/hr or less depending on delivery rate) or Continue Rate (CR) to continue at the current rate
  • Variable distal pressure setting
  • Nonpulsatile volumetric accuracy
  • Microprocessor control
  • Large LCD
  • Panel back illumination on mains power
  • Lockout switch
  • Standard fulfill, partfill, syringe, and vial use
  • Enteral and parenteral fluid delivery
  • Blood and blood product delivery

phone Call for price: (855) 477-8866

Download the operator's Manual

Description

Here are the specifications of Hospira Plum A+ with Mednet Infusion Pump Repair.

Physical:
  • Dimensions: 8” H x 8” W x 6” D (excluding pole clamp and power cord storage)
  • Weight: 9.5 lbs (with battery)
  • Casing: High-impact plastic
Electrical:
  • Power Requirements: 120 VAC; 50-60 Hz; 35 W
  • Power Cord: Hospital-grade AC cord; 10 feet; with transparent plug and retainer
  • Fuses: 0.5 A, 250 VAC
  • Battery: Sealed lead acid; 6 V; internal; rechargeable
  • Battery Operation: A fully charged new battery provides approximately three hours of operation at 125 mL/hr, or delivers 250 mL if > 126 mL/hr
  • Recharge Time: Less than 6 hours (if operating at 125 mL/hr on one line)
  • The battery charges whenever the infuser is connected to AC power

Self-Discharge: 50% of charge is retained for a minimum of one month when the infuser is not connected to AC power or is not operating

Nurse Call System: Default: Normally-open (NO)

Delivery Rate Ranges:
  • Delivery Rate (Lines A and B): 0.1 to 99.9 mL/hr (in 0.1 mL/hr increments); 100 to 999 mL/hr (in 1 mL/hr increments)
  • Concurrent Delivery: 0.5 mL/hr minimum for each line
  • PlumSet: 500 mL/hr cumulative (A + B) maximum
  • KVO: 1 mL/hr or the last primary delivery rate, whichever is less
  • VTBI Range: 0.1 to 99.9 mL (in 0.1 mL/hr increments); 100 to 9999 mL (in 1 mL/hr increments)
  • Accuracy: ± 5%
Alarms and Limits:
  • Distal Occlusion: The distal occlusion alarm sounds after the distal tubing or set outlet fitting becomes occluded
  • Proximal Occlusion: The proximal occlusion alarm sounds within two pumping cycles when the tubing proximal to the cassette becomes occluded
  • Distal Pressure Limit
    (without alarm): 1 to 15 psi; maximum pressure limit is user-selectable; factory setting is 6 psi
  • Maximum Infusion Pressure: 20 psi

Air-in-Line Alarm:

    • PlumSet (Distal): Bolus – 0.5 mL of air or larger; Cumulative – 0.5 mL of air out of 5.3 mL of fluid
    • PlumSet (Proximal): Bolus at 0.5 mL, total 1 mL (0.5 mL concurrent)
Communication:

Ethernet LAN: Shielded Ethernet cable plugged into an RJ-45 connector

Wireless LAN: Device name: Hospira MedNet Wireless 802.11 a/b/g Module

    • Standards: IEEE 802.11 a/b/g
    • Transmit Power: 802.11 b/g – 17.6 dBm; 802.11 a – 19 dBm
    • Antenna: Integrated Surface Mount Antenna
    • Certification: FCC Part 15.247, 15.407; IC RSS-210, RSS-102
    • FCC ID: STJ-80411396001, IC: 5627A-80411396

Ethernet IP Address: 192.168.0.100

Subnet Mask: 255.255.0.0

Alarms
  • Distal Occlusion
  • Proximal Occlusion
  • Distal air-in-line
  • Proximal air-in-line
  • Low battery
  • Door open while pumping
  • Lockout violation
  • VTBI complete
  • Valve/cassette test failure
  • Nurse call
  • No action alarm
  • Infuser idle for two minutes

Download the operator’s Manual

https://web.facebook.com/infusionpumprepair/photos/a.101426038761773/102807338623643

Recharge Time: Less than 6 hours (if operating at 125 mL/hr on one line)

    • The battery charges whenever the infuser is connected to AC power

Self-Discharge: 50% of charge is retained for a minimum of one month when the infuser is not connected to AC power or is not operating