- BD McGaw 360 Infuser -

BD McGaw 360 Infuser

BD McGaw 360 Infuser

The McGaw BD 360 syringe pump features continuous intravenous infusion rates up to 360 mL/hr, depending upon syringe size. The BD 360 can be used as a primary infusor or can deliver downstream secondary piggyback infusions. Variable time settings are set using digital controls, and the LCD display shows time remaining.

phone Call for price: (855) 477-8866

Download the operator's Manual

Description

BD McGaw 360 Infuser

So here you will be able to find the dimensions, weight, syringes and rate parameters of BD McGaw 360 Infuser Infusion Pump Repair .

  • Dimensions: 9.75” H x 2.75” W x 5” D
  • Weight: 2 lbs
  • Syringes: BD Plastipak™ 3cc to 60cc
  • Rate Parameters: 1.5 mL/hr (3cc syringe) to 360 mL/hr (60cc syringe)
  • Delivery Time: 10 – 60 min., in 2.5 min. increments
  • Battery: Requires 4 C-size alkaline batteries
  • Battery Life: 400 hours typical, 10 hours on Low Battery
  • Accuracy: ± 3%

Displays:

    • LED: Infuse, Attention, Low Battery, Infusion Complete.
    • LCD: Time Remaining, Original Set Time; some revisions have Infusion complete indicator.

Occlusion Limits (using BD Plastipak™ syringes):

Syringe

Size

Maximum Occlusion

Pressure (psi)

Typical Occlusion

Pressure (psi)

3 cc 46.8 36.4
5 cc 30.0 17.9
10 cc 22.8 17.0
20 cc 21.6 13.9
30 cc 16.8 9.4
60 cc 13.2 8.4

BD McGaw 360 Infuser

Here are some features of BD McGaw 360 Infuser.

Dimensions: 9.75” H x 2.75” W x 5” D

Weight: 2 lbs

Syringes: BD Plastipak™ 3cc to 60cc

Rate Parameters: 1.5 mL/hr (3cc syringe) to 360 mL/hr (60cc syringe)

Delivery Time: 10 – 60 min., in 2.5 min. increments

Battery: Requires 4 C-size alkaline batteries

Battery Life: 400 hours typical, 10 hours on Low Battery

Accuracy: ± 3%

Displays:

    • LED: Infuse, Attention, Low Battery, Infusion Complete.
    • LCD: Time Remaining, Original Set Time; some revisions have Infusion complete indicator.

Occlusion Limits (using BD Plastipak™ syringes):

Syringe

Size

Maximum Occlusion

Pressure (psi)

Typical Occlusion

Pressure (psi)

3 cc 46.8 36.4
5 cc 30.0 17.9
10 cc 22.8 17.0
20 cc 21.6 13.9
30 cc 16.8 9.4
60 cc 13.2 8.4

 

Dimensions: 9.75” H x 2.75” W x 5” D

Weight: 2 lbs

Syringes: BD Plastipak™ 3cc to 60cc

Rate Parameters: 1.5 mL/hr (3cc syringe) to 360 mL/hr (60cc syringe)

Delivery Time: 10 – 60 min., in 2.5 min. increments

Battery: Requires 4 C-size alkaline batteries

Battery Life: 400 hours typical, 10 hours on Low Battery

Accuracy: ± 3%

Displays:

    • LED: Infuse, Attention, Low Battery, Infusion Complete.
    • LCD: Time Remaining, Original Set Time; some revisions have Infusion complete indicator.

Related Keyphrase:

  • BD McGaw 360 Infuser Infusion Pump

Download the operator’s Manual

https://web.facebook.com/infusionpumprepair/photos/a.101426038761773/102015452036165