- Kangaroo 924 -

Kangaroo 924

Kangaroo 924

Kangaroo 924 Enteral Feeding Pump The Kangaroo™ 924 Enteral Feeding Pump features pre-set volume and overflow infusion safeguards to ensure the correct and consistent delivery of nutrition. The easy-to-use controls make set-up and operation of the unit intuitive.

Features
  • Portable operation-built in battery
  • Wide flow rate range:

-1-5ml/hr in 1ml/hr increments

-50-300 ml/hr in 5 ml/hr increments

  • Small and lightweight for easy transport
  • Touch panel operation
  • Shut off and activation of audible and visible alarms when:

-feeding container is empty

-feeding tube becomes occluded

-drop sensors blocked

-battery low (alarms 15 minutes before pump shut off)

-unit is left on “HOLD” longer than 2 1/2 minutes

-pump set is improperly loaded

-self diagnosed problem which requires technical servicing

-pump motor malfunction

  • Automatic systems check after pump has been turned ON
  • Large LED display
  • Removable power cord
  • Low occlusion pressure. Pump will not overcome back pressure greater than approximately 12psi (82.7kPa)
  • Integral pole clamp
  • Use of safety interlock pump sets to prevent free flow from inadvertent misload of pump set onto the pump
  • Pump running indicator lights
  • DC motor with intermittent operation
  • Adjustable alarm volume
  • Memory of flow rate, dose, and volume delivered settings for 24 hours after pump shut off

phone Call for price: (855) 477-8866

Download the operator's Manual

Description

Kangaroo 924

So here you will be able to find the specifications of this  Infusion Pump Repair and kangaroo 924 service manual. Here are the Type Infusion Device, pumping mechanism, feeding set required, operating range , accuracy , alarms , battery, dimensions, weight, power requirements and etc. Download kangaroo 924 enteral feeding pump

  • Type Infusion Device: Volumetric
  • Pumping Mechanism: Rotary Peristaltic
  • Feeding Set Required: Covidien Kangaroo 924 Enteral Feeding Pump with Pole Clamp Feeding Sets (see www.covidien.com for details on compatible administration sets)

Operating Range:

  • Rate: 1-300 mL/hr in 1 mL/hr increments
  • Dose: 1-2000 mL in 1 mL increments
  • Volume Delivered: 1-9999 mL (rollover is accounted for)

Accuracy: 10% or 0.5 mL/hr, whichever is larger, with top of the fluid column at 71.12 cm (28”) ± 0.76 cm (0.3”) above the top of the pump, at a room temperature of 22 °C ± 2 °C (72 °F ± 3 °F), using a new 924 feeding set for no longer than the recommended 24 hours of maximum usage.

Occlusion Pressure: 15 psi (103.4 kPa) Nominal

Alarms:
  • Flow Error • Low Battery • Hold Error • Load Set • System Error • VTBD Delivered
Battery:
  • Rechargeable NiMH
  • Recharge time for fully discharged battery is approximately 12 hours.
  • A new battery, when fully charged, will operate for approximately 16 hours at 125 milliliters per hour.

Dimensions: 7.25” High x 6.61” Wide x 4.33” Deep

Weight: Approximately 3.89 lbs (1.76kg)

Case Material: ABS/PC Plastic – Fire Resistant

Power Requirements: 120V~, 60 Hz, 1 Amp (North America); 230V~, 50Hz, 0.5 Amp (UK / Australia). If the integrity of the protective earth conductor arrangement is in doubt, operate unit from its internal electrical power source.

Ambient Operating Conditions: 10° to 40°C (50° to 105°F)

Transport and Storage (not to exceed 30 days): 0° to 50°C (32° to 122°F),10-95% RH, non-condensing.

Kangaroo 924

Download kangaroo 924 enteral feeding pump

Type Infusion Device: Volumetric

Pumping Mechanism: Rotary Peristaltic

Feeding Set Required: Covidien Kangaroo 924 Enteral Feeding Pump with Pole Clamp Feeding Sets (see www.covidien.com for details on compatible administration sets)

Related Keyphrase:

  •  Infusion Pump Repair
  •  kangaroo 924 service manual

Download the operator’s Manual