- CADD Legacy Plus -

CADD Legacy Plus

CADD Legacy Plus

The CADD-Legacy™ PLUS ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-Legacy™ PLUS pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous or intermittent rate of infusion.

Features
  • Easy to read display
  • Cassette detection
  • Occlusion detection
  • Reliable and Durable
  • Programming flexibility
  • Air-in-Line detector with Off, High- and Low-sensitivity settings
  • ON/OFF key allows pump to be put in a low-power state to save battery life
  • Event memory of the last 1,000 events in the pump’s history

Categories: ,

phone Call for price: (855) 477-8866

Download the operator's Manual

Description

CADD Legacy Plus

So here you will be able to find the specifications of CADD Legacy Plus Infusion Pump Repair and CADD Legacy plus pump manual. Here are the dimensions, weight, moisture protection, power and etc. Download cadd legacy plus error codes

  • Dimensions: 1.6″ W X 3.8″ H X 4.4″ L

  • Weight: 0.86 lbs. (13.8 oz), including 2 AA batteries and empty medication reservoir
  • Moisture Protection: Splash-proof (IPX4)
  • Power: Two AA alkaline batteries; AC Adapter
  • Battery Capacity: The expected life of 2 AA batteries is 15 hours at 125 ml/hour, or approximately 14 days at 10 ml/day (nominal).
  • Accuracy: ± 6% (nominal)
  • Maximum Infusion Pressure: 40.0 psi
  • Maximum Time to Occlusion Alarm: CADD® Administration Set with anti-siphon valve: 2.0 hours
  • Operating Temperature: +35°F to 104°F (2°C to 40°C)
  • Storage/Transportation Temperature: -4°F to 140°F (-20°C to 60°C)
  • High Pressure Alarm: 26 ± 14 psi
  •  Maximum Volume Infused under Single Fault Condition: CADD® Administration Set with anti-siphon valve: 0.2 ml
  • Delivery Rate during priming: Approximately 180 ml/hr
  • Alarm Disabled during priming:  Air Detector
Continuous Delivery Specifications:
  • Reservoir Volume: 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default = 1.0 ml)
  • Continuous Rate:1 to 125 ml/hr; programmable in 0.1 ml/hr increments (Default = 0.0 ml)
  • Given: 0 to 99999.95 in 0.05 ml increments
Intermittent Delivery Specifications:
  • Reservoir Volume: 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default = 1.0 ml)
  • Dose Volume: 1 to 1000 ml programmable in 0.1 ml increments (Default = 0.0 ml)
  • Dose Duration: Default = 30 min. Duration lasts 1 min to 24 hrs in the following increments:
    • 1 minute for values between 1 min and 10 min
    • 5 minutes for values above 10 min
    • * Duration is limited by Dose Volume so that rate does not exceed 125 ml/hr.
  • Dose Cycle: 10 min to 96 hrs in 5 minute increments. Default = 4 hrs
  • KVO Rate: 0 to 125 ml/hr in 0.1 ml/hr increments. Default = 0 ml/hr
  • Given: 0 to 99999.95 in 0.05 ml increments
  • Dose Remaining: 1 minute increments

Alarms/Alerts:

Low battery power Disposable not attached when run attempted
Depleted battery power Motor locked
Battery dislodged Upstream occlusion
Pump stopped Reservoir volume empty
Pump fault Program incomplete
Low reservoir volume Key stuck
High delivery pressure Disposable detached
Air in line

CADD Legacy Plus

Download cadd legacy plus error codes

  • Dimensions: 1.6″ W X 3.8″ H X 4.4″ L
  • Moisture Protection: Splash-proof (IPX4)
  • Power: Two AA alkaline batteries; AC Adapter