- CADD Legacy -

CADD Legacy

CADD Legacy

The CADD-Legacy™ 1 ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-Legacy™ 1 pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. The pump is intended for therapies that require a continuous rate of infusion.

Features
  • Easy to read display
  • Cassette detection
  • Occlusion detection
  • Air-in-Line detector with Off, High- and Low-sensitivity settings
  • ON/OFF key
  • Event memory
  • Reliability
  • Programming flexibility

phone Call for price: (855) 477-8866

Download the operator's Manual

Description

CADD Legacy

Here are the specifications of CADD Legacy Infusion Pump Repair and CADD legacy pump service manual. Download CADD legacy cassette

  • Dimensions: 1.6″ W X 3.8″ H X 4.4″ L
  • Weight: 0.86 lbs. (13.8 oz), including 2 AA batteries and empty medication reservoir
  • Resolution: Medication Cassette Reservoir or CADD Administration Set, 0.050 ml per pump stroke nominal
  • Moisture Protection: Splash-proof (IPX4)
  • Power: Two AA alkaline batteries; AC Adapter
  • Battery Capacity: The expected life of 2 AA batteries is 15 hours at 125 ml/hour, or approximately 14 days at 10 ml/day (nominal). It is recommended that 2 new AA batteries be kept available for replacement if necessary.
  • Accuracy: ± 6% (nominal). At low infusion rates, this accuracy may not be achieved for short periods.
  • Maximum Infusion Pressure: 40.0 psi
  • Maximum Time to Occlusion Alarm: CADD® Administration Set with anti-siphon valve: 2.0 hours
  • Bolus Volume at Occlusion Alarm Pressure: 0.050 ml resolution CADD® Administration Sets/Medication Cassette Reservoirs: <0.15 ml
  • Operating Temperature: +35°F to 104°F (2°C to 40°C)
  • Storage/Transportation Temperature: -4°F to 140°F (-20°C to 60°C)
  • High Pressure Alarm: 26 ± 14 psi
  • Air Detector Alarm:
    • Single bubble Low Sensitivity: > 0.250 ml
    • Single bubble High Sensitivity: > 0.100 ml
    • Multi-bubble: 1.0 ml nominal
  •  Maximum Volume Infused under Single Fault Condition: CADD® Administration Set with anti-siphon valve: 0.2 ml
  • Delivery Rate during priming: Approximately 180 ml/hr
  • Alarm Disabled during priming:  Air Detector
  • Continuous Delivery Specifications:
    • Reservoir Volume: 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default = 1.0 ml)
    • Continuous Rate: 1 to 3000 ml/24 hr; programmable in 1 ml/24 hr increments (Default = 1.0 ml)
    • 0 to 99999.95 in 0.05 ml increments
Alarms/Alerts
  • Low battery power
  • Depleted battery power
  • Battery dislodged
  • Pump stopped
  • Pump fault
  • Low reservoir volume
  • High delivery pressure
  • Air in line
  • Motor locked
  • Upstream occlusion
  • Reservoir volume empty
  • Key stuck
  • Disposable detached

Download the operator’s Manual

https://web.facebook.com/infusionpumprepair/photos/a.101426038761773/102794281958282

Related Keyphrase:

  • Infusion Pump Repair
  • cadd legacy pump service manual
  • cadd legacy cassette