CADD Prizm VIP

CADD Prizm VIP

The CADD-Prizm® ambulatory infusion pump provides measured medication therapy to patients in hospital or outpatient settings. The CADD-Prizm® pump is indicated for intravenous, intra-arterial, subcutaneous, intraperitoneal, epidural space, or subarachnoid space infusion. PCA delivery is intended for therapies that require a continuous rate of infusion, patient-controlled demand doses, or both (such as patient-controlled analgesia). Continuous delivery allows the infusion of drug at a constant, programmed rate while intermittent delivery allows the infusion of a specific volume of drug at a regular, programmed interval. The final delivery program, TPN (Total Parenteral Nutrition), allows the infusion of nutritional solutions or other fluids, with optional tapering at the beginning and end of infusion.

Features
  • Easy to read display
  • Reliable and Durable
  • Programming flexibility
  • Air-in-Line detector
  • Event memory log
  • Lock levels for patient safety

Categories: ,

phone Call for price: (855) 477-8866

Download the operator's Manual

Description

CADD Prizm VIP

Here are the specifications of  CADD Prizm VIP pump manual. Download cadd prizm vip 6101

  • Dimensions: 1.7″ W X 4.1″ H X 5.6″ L
  • Weight: 1.25 lbs. (20 oz), including 9 volt battery and empty medication reservoir
  • Resolution: Medication Cassette Reservoir or CADD Administration Set, 0.050 ml per pump stroke nominal
  • Power: 9 volt alkaline or lithium battery such as DURA-CELL® Alkaline MN 1604 or ULTRALIFE® Lithium U9VL; CADD External Power Source (EPS) Power Pack (reorder number 21-3801); AC Adapter.
  • Battery Capacity: The expected life of a 9 volt battery is 12 hours at 100 ml/hour, or approximately 5 days at 10 ml/day (nominal). It is recommended that a new 9 volt battery be kept available for replacement if necessary.
  • System Operating Temperature: +36°F to 104°F (2°C to 40°C)
  • System Storage Temperature: -4°F to 140°F (-20°C to 60°C)
  • Power Pack Charging Temperature: +50°F to 95°F (10°C to 35°C)
  • Accuracy: ± 6% (nominal)
  • Bolus Volume at Occlusion Alarm Pressure: 0.050 ml resolution administration Sets/Medication Cassette Reservoirs: <0.25 ml; 0.100 ml resolution administration sets: <2.0 ml
  • High Pressure Alarm: 18 ± 9 psi
  • Air Detector Alarm: Single bubble greater than 0.100 ml (can be turned ON or OFF)
PCA Delivery Mode :
  • Reservoir Volume: 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default: 1 ml)
  • Units: milliliters (ml), milligrams (mg), micrograms (mcg); (Default: milligrams)
  • Concentration:
    • Mg/ml: 0.1, 0.2, 0.3, 0.4, 0.5, 1, 2, 3, 4, 5, 10, 15, …95, 100
    • Mcg/ml: 1, 2, 3, 4, 5, 10, 15, …95, 100, 200, 300, 400, 500
    • (Default = 1)
  • Continuous Rate: 0-30 ml/hr (or the mg or mcg equilavent); Default: 0 mg/hr
  • Demand Dose:
    • 0 to 9.9 ml (Default: 0 mg/hr)
    • Delivery rate (Continuous Rate + Demand Dose): 125 ml/hr nominal
  • Demand Dose Lockout: 5 minutes to 24 hours in the following increments:
    • 1 minute for values between 5 and 20 minutes
    • 5 minutes for values between 20 minutes and 24 hours
    • (Default: 5 minutes)
  • Max Doses per Hour: 1 – 12 doses in 1 dose increments (will also be limited by the Demand Dose Lockout value); Default: 1
  • Demand Doses Given: 0 to 999
  • Demand Dose Attempts: 0 to 999
  • Given: 0 to 99999.99 in 0.01 unit increments
  • Clinician Bolus:
    • 0.1 ml to 20.00 ml (or mg or mcg equivalent)
    • Delivery rate (Continuous Rate + Clinician Bolus): 125 ml/hr nominal
Continuous Delivery Specifications:
  • Reservoir Volume: 1 to 9999 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments Default: 1 ml
  • Continuous Rate: 0.1 – 350 ml/hr in the following increments:
    • 1 for values between 0.1 and 100
    • 1 for values between 100 and 350
    • Default: 0 ml/hr

*Use 9 volt battery for rates up to 250 ml/hr; use power pack or AC adapter for rates up to 350 ml/hr.

  • Given: 0 to 99999.9 in 0.1 unit increments

TPN Delivery Mode Specifications:

  • Reservoir Volume: 0 to 9990 or Not In Use; programmable in 1 ml increments, displayed in 0.1 ml increments (Default: 10 ml)
  • Infusion Volume: 10 – 9990 ml in 10 ml increments (Default: 10 ml)

Download the operator’s Manual